The 97-Well Rubber Pad has precisely distributed wells and is designed for the corresponding 97-well quantitative detection tray. It is resistant to high and low temperatures.
Product specifications
- Power supply: 220 V ±10%, 50 Hz
- Ambient temperature: −10°C to 50°C
- Relative humidity: 20%–90%, without heavy dust or corrosive gas
- Noise: ≤50 dBA
- Cover temperature: ≤40°C
- Preheating time: ≤30 minutes
- Weight: 1 kg
- 97-well tray sealing time: less than 15 seconds
- Detection range: 0–2419 MPN/100 mL for undiluted water samples
Use
Preheat the sealer until the green indicator confirms operating temperature. Add the coliform enzyme-substrate reagent to a 100 mL water sample and mix gently. Pour the sample into the 97-well tray, remove bubbles, and align the tray with the corresponding wells on the rubber pad. Keep the tray openings facing upward, then feed the assembly into the sealer. The tray and pad are released after sealing.
Incubate total coliform and E. coli samples at 36 ±1°C. For thermotolerant coliform, incubate at 44.5 ±0.2°C for 24 ±2 hours. Read positive wells and calculate the result using the applicable MPN table and dilution factor.